ASTM D6730 DANI Optimized Method with Cryo
Oven
- Initial temperature 5 °C
- Initial time 10 min
- First program rate 5 °C/min
- First hold temperature 50 °C
- First hold time 0 min
- Second program rate 1.5 °C/min
- Final temperature 200 °C
- Final hold time 5 min
Injector
- Temperature 250 °C
- Split ratio 150:1
- Sample size 0.5 uL
Detector
- FID
- Temperature 250 °C
- Hydrogen 30 mL/min
- Air 300 mL/min
- Nitrogen 30 mL/min
Carrier gas
- Helium, pressure constant: 4.16 bar