ASTM D6730
Oven
- Linear carrier flow (He) 2.4 ml/min
- Initial temperature 35 °C
- First hold temperature 5 °C
- First hold time 10 min
- First program rate 5 °C/min
- Second temperature ≈ 49
- Second hold time ≈ 55
- Second program rate 1.6 °C/min
- Final temperature 200 °C
- Final hold time 5 min
- Analysis time ≈ 170
Injector
- Temperature 250 °C
- Split ratio 100:1
- Sample size 0.1 - 0.2 uL
Detector
- FID
- Temperature 250 °C
- Hydrogen flow 35 mL/min
- Air flow 350 mL/min
- Make up flow He 20 mL/min
Carrier gas
- Helium
- EPC pressure 301 kPa
- Linear velocity 28 cm/s
- Gas saver flow 25
- Gas saver time 2