SST Properties - Calculate By
Tabs: General - Parameters - Subparameters - Calculate By - Events
The method of an individual parameter calculation is selected on the SST Properties - Calculate By tab. The selection of the calculation method effects the following parameters: efficiency (Efficiency), length efficiency (Efficiency/Length), height equivalent of theoretical plate (HETP) and resolution (Resolution).
Note:
A mathematical formula for parameter calculation according to selected calculation methods are found in the chapter "Appendix - Mathematical Formulas"
EP/USP (European Pharmacopoeia, United States Pharmacopoeia since 12/2022, also DAB, BP, ASTM)
Calculation of efficiency and related parameters based on the width at half height.
USP obsolete (United States Pharmacopoeia until 12/2022)
Calculation of efficiency and related parameters based on width given by intersection of tangents with baselines.
JP (Japanese Pharmacopoeia)
Calculation of efficiency and related parameters based on the width at half height.