This usually happens when upgrading Clarity from an earlier version. The IQ program checks all registered files and, if they are no longer used in the new version, will cause the check to fail. However, the station will usually function correctly.
To get Passed for IQ results, it is necessary to:
- Remove the previous Clarity installation using the Uninstall Clarity command from the Windows Start menu - All Programs - Clarity program group.
- In case you can still see some *.dll files in C:\Clarity\Bin (since 7.0), delete them manually.
- Install Clarity again.
The Uninstall Clarity command will remove the program files only, leaving the configuration and data intact.
In Clarity versions prior to 7.0, *.dll files were present in the C:\Clarity root directory.