Clarity sequence does not progress when started or running.
There may be multiple different reasons. Please check the detailed status in the Instrument window and Device Monitor.
- Waiting for Injection - indicates parameters for injection were sent to the sampler, but the start signal has not yet been received.
Suppose the sampler is not working on the injection. In that case, it may indicate wrong synchronization settings in the System Configuration or that the start signal was not received or ignored for some reason. - Waiting for Hardware to be Ready to Run - indicates some of the devices configured on the instrument do not report Ready state.
Check the status reported by individual devices in the Device Monitor. - Sequence Paused - indicates the sequence was paused.
An automatic pause may happen if an invalid value is entered during the editing of the running sequence to prevent locking a line with an invalid value. After correcting the value, the sequence must be manually resumed.