I have a Gow-Mac GC chromatograph. How should I connect it with Clarity station?

For the Gow-Mac GC chromatograph, use the asymmetrical connection as shown in the picture below.

Connect Brown wire (pink tubular end sleeve) to terminal (+) and white wire (grey tubular end sleeve) together with yellow wire (shielding) to terminal (-).

Do not connect anything to the terminal marked D in the picture, it may cause damage to the A/D converter. Set also the Signal Range at the Method Setup Acquisition tab to the lowest possible value.

gow-mac-wiring