5.0.3

*New: Control: Advion Expression CMS MS detector released.
*New: Control: Eksigent - UltraLC 100/110 - HPLC System has been released.
*New: Control: Spark Holland Symbiosis control module (Alias, ACE, HPD) released.
*New: Calibration: Named Groups defined in the Calibration will be displayed in the Chromatogram - Result table below the Total Row. One compound could be present in several named groups. Not supported in the Chromatogram - Summary table.
*New: Calibration: New response base Area% - for special calculations in clinical diagnosis.
*New: Clarity2Go: Option to set password for the Clarity2Go utility.
*New: Control: Agilent 7890 implemented check for no data from detectors (problem on computers with multiple LAN boards).
*New: Control: Antec Decade Elite - Find button in the Setup dialog to detect available instruments on network.
*New: Control: Dani Master MS-TOF detector driver available for testing.
*New: Control: Knauer control module drivers updated to version 5.0.1.4095.
*New: Control: Rigol L-3000 driver (version 1.0.0.28) is now available for testing.
*New: Control: Rigol L-3320 As driver updated to version 1.0.0.19.
*New: Control: SSI - Mighty-Mini - pump is now available for testing.
*New: Control: Sykam S1130, S1132 pump driver updated to version 1.0.4.2 and S3250 driver updated to version 1.0.5.2.
*New: Control: Techlab Minipump 1 control driver by Uni-Ruby script available for testing.
*New: Control: Watrex - Mighty-Mini - pump is now available for testing.
*New: Control: YL Instruments YL 6500 GC driver updated to version 1.0.1.4.
*New: Extension CE: Calculation of Apparent Mobility and Effective Mobility is now available in the Method Setup - Performance tab.
*New: Extension GPC: The Method Setup, GPC Advanced tab has been removed.
*New: Extension PDA: The Remove Generated Signal option enables to remove signals generated from PDA view.
*New: Chromatogram: Create Label option Active Chromatogram renamed to Active Signal.
*New: Chromatogram: Detector delay is now enabled also on the first signal.
#Fixed Bug: Calibration: Groups could not be evaluated by ISTD (introduced in 5.0.0 with multiple ISTD).
#Fixed Bug: Calibration: Peak coloring did not function correctly in Calibration window for Groups.
#Fixed Bug: Control: Advion CMS detector contact closure output not working.
#Fixed Bug: Control: Agilent 1200 ICF HPLC - possible missed starts on system without 1200 HPLC system detector.
#Fixed Bug: Control: Agilent 7890 GC -  oven max temperature limit check has been set to 70-450 °C.
#Fixed Bug: Control: Agilent 7890GC/7683AS wrong syringe sizes offered in the Agilent 7890 Setup - Ampler tab.
#Fixed Bug: Control: Decade Elite Temperature range changed to 15 - 60 °C.
#Fixed Bug: Control: Knauer PDA-1 detector - added error message when the spectral data could not be processed due to low computer memory.
#Fixed Bug: Control: Sedere Sedex LC - possible freeze during acquisition.
#Fixed Bug: Control: Shimadzu GC 2014,  Gas Saver settings, interchanged values for Split Ratio and Time.
#Fixed Bug: Control: Sykam S3210/40 DAD detector: Wrong signal axis units for displaying spectra.
#Fixed Bug: Control: UNI-RUBY scripts - non functional Time Tables (introduced in version 4.0).
#Fixed Bug: Control: UNI-RUBY scripts: Ruby Exception when a Bypass injecton was used.
#Fixed Bug: Error message "DEMO - Wrong Serial Number" changed to more appropriate "DEMO-Wrong User Code".
#Fixed Bug: Export AIA from Postrun exported only the C1 completeness. Now it exports C1 + C2.
#Fixed Bug: Chromatogram: User columns defined in the Summary table are not displayed (already fixed in 5.0.2).
#Fixed Bug: Possible crash when default printer was not configured in Windows.
#Fixed Bug: Print to PDF: New version of the DynaPDF library 3.0.  Should solve the problems in landscape printing.
#Fixed Bug: Print to PDF: the possibility to print Chromatogram to multiple pages in landscape mode.
KNOWN ISSUE: The printing of reports to PDF does not print the graphs properly. Suggested workaround: use the free PDF Creator to print reports to PDF.

Product Line
Product Version
5.0
Released