*New: Calibration: new option "Show recalibration points" was added into Graph Properties to display recalibration points in a calibration curve graph.
*New: Chromatogram: new field “Chromatogram Comments” was added to the Measurement tab in the Chromatogram window. It serves for notes concerning the chromatogram and its content can be printed in the report below the Result tables. This field is stored within the chromatogram file.
Report Setup: new Chromatogram Comments checkbox for printing of the comments was added to the Report Setup – Results tab.
*New: Clarity DEMO: new demo data, the demo now better illustrates the suggested workflow and available features.
*New: AS Control - Spark Holland Midas and Marathon: the status of these Spark autosamplers is now displayed in the Device Monitor.
*New: LC Control: Agilent 1100 pump state is now displayed in the Device Monitor.
*New: AS Control - HTA samplers: the status of HTA samplers is now displayed in the Device Monitor.
*New: Help: the help files format was changed from *.hlp to *.chm for compatibility with Windows Vista.
*New: Audit trail: triggering the Stop Flow button/command from the Device monitor is now recorded in the Audit Trail.
*New: Chromatogram - Results table: the parameters from Performance table are now available as columns in the Result table (they are hidden by default).
*New: Single Analysis: the time variables in the chromatogram name (%T, %R etc.) reflect the start of analysis instead of end in Single run mode to unify the behaviour with a Sequence mode.
*New: AS Control: new CTC PAL driver is now able to control larger number of CTC autosampler types (including former CTC GC PAL).
*New: Audit trail: changes to the chromatogram header entries (Sample ID, Sample, Amount, Inj. Volume, ISTD Amount, Dilution) are recorded in the Station or Chromatogram Audit trails (when the corresponding Intergration Table checkbox in in the Audit Trail Settings - Chromatogram dialog is checked).
*New: Sequence: chromatograms can now be stored in subfolders of the project data directory, the subfolder must already exist before use.
*New: Data Acquisition: new option to set display ranges individually for different signals when the "Common for All Signals" checkbox is unchecked was added to the Data Acquisition window.
*New: Chromatogram: an option to define units for user defined columns in the Result and Summary tables was added.
*New: Data Acquisition: auxiliary signals can now be displayed in the graph of the Data Acquisition window.
*New: Configuration: new device type Auxiliary (reserved for future use) was added to the Setup Control Modules list.
*New: Clarity: New "Options" dialog (accessed from Clarity window using the System - Options... command) contains "Enable chromatogram files overwriting" checkbox. When unchecked, the Save and Save As commands will not allow to overwrite existing chromatogram file. Automatic saving will in such case modify the filename to make it unique.
*New: Report Setup - Calibration dialog: new "Calibration history" checkbox enables to print the history of recalibrations - the report can include Compounds and Levels, both either Current, Valid or All.
*New: LC Control - Fraction Collector: in the Data Acquisition window, the actual Fraction delay for the FC GP control module is displayed by vertical line.
*New: Data Acquisition: new Overlay toolbar was added to the Data Acquisition window to enable signal switching. When updating from previous versions of Clarity, the Reset All command from context menu has to be performed to display it.
*New: Directories dialog: when a new directory is created, files from the default Common folder will be copied to the Common folder in the new structure.
*New: Hardware key: RkNDUSB (Rockey 4 ND) hardware key is newly supported
*New: Control: when a not communicating device is detected during instrument opening, the communication with other devices will not be tested.
*New: AS control - CTC-PAL: new option Prep. Ahead was added to the Setup Configuration dialog enabling a prepartion of next sample during the previous run.
*New: AS Control - Spark Holland Midas, Endurance and Triathlon samplers: new option "Sequence Mode" is added into the Spark Autosampler Setup dialog. It enables correct performing of the washes set for injection/vial/step. When not checked, each injection is sent to the sampler as individual step.
*New: Command line: command line parameter prm_import_ex creates the name of the imported chromatogram from the name of the source file.
*New: LC Control: new Shimadzu LC-10/20 system control module (requiring the system controler) is now available.
*New: AS Control: Spark Holland Endurance control module was replaced with a new one.
*New: LC Control: detector control module for a new device - Antec Leyden Decade II - is now available.
*New: LC Control: Bischoff 2250 pump module was replaced with a new version.
#Fixed Bug: Chromatogram - Export: the export from the Result table now correctly exports User defined columns when not in the "In Fixed Format" mode.
#Fixed Bug: Chromatogram - Overlay: the signal range for mathematical operations (move) was increased from 10000 to 100000 mV (or default units respectively).
#Fixed Bug: Print report: the bug causing the Logarithmic axes setting to be ignored when printing from Sequence window was repaired.
#Fixed Bug: LC Control: it is now possible to configure up to 4 Agilent 1100 systems using LAN control on Clarity station.
#Fixed Bug: Print report: Noise and Drift parametres of the Result table were not printed in Clarity version 2.5.6.
#Fixed Bug: Report Setup - Page setup: it is now possible to set the Header font, including the font color.
#Fixed Bug: Extensions - GPC: the units displayed for K (dL/g *10^3) were corrected to dL/g *10^5.
#Fixed Bug: AS Control - CTC A200S: incorrect sending of certain injection parameters was amended.
#Fixed Bug: Print Report - Method: integration table parameters incorrect formating in the report was repaired.
#Fixed Bug: LC Control - gradient table: Gradient table authomatically sorts itself by time after time value change was performed on any row.
#Fixed Bug: GC Control: updating of values during the analysis in the Device Monitor for DANI 1000 GC.
#Fixed Bug: Report Setup: the Save button was removed. When the report is modified and the user hits Cancel button, Clarity will prompt for saving the changes. Changes are saved automatically after pressing OK.
#Fixed Bug: Calibration: the "Add Existing" command in Calibration mode now correctly increments the current level.
#Fixed Bug: Control: The Ready Out signal worked incorrectly in case a digital output was activated on the same device at time 0 from the Event table.