*New: All command line parameters received are now recorded in Audit Trail.
*New: German localization is now available for everyone free of charge.
*New: TeamviewerQS utility for remote support was added to the installation, available in Start menu, Program Files, Clarity Lite group.
*New: View Calibration command in Chromatogram window will open both the active chromatogram and the linked calibration in the Calibration window.
#Fixed Bug: Changed method in sequence running in Sequence mode was not properly propagated to instrument.
#Fixed Bug: injection_date_time_stamp and sample_id_comments fields are now supported during the Import CDF function.
#Fixed Bug: Mathematical operations creating multisignal chromatograms will now set signal names.
#Fixed Bug: Possible crash using parenthesis after %T variable in filenames.
#Fixed Bug: Prints are now performed in separate thread (should solve some communication errors due to timeouts caused by blocking the main thread).
#Fixed Bug: Set Calibration command failed for calibration files using dots within the file name on the Method Setup - Calculation tab.