*New: The Clarity software is compatible with Windows 10.
*New: Chromatogram - Baseline Crossing includes a new option to integrate the peak area "To Start/End" in addition to existing "To First Cross" option.
*New: Chromatogram - Column Amount [%] displaying the percentages calculated from Amount [user units] column was renamed to Amount% to distinguish them when user units were set to %.
*New: Chromatogram - Event table the output command Hold/Resume for gradient control (supported only by selected pump models).
*New: Chromatogram - Support for ISTD1- ISTD5 in the Result Table in the User Columns.
*New: Chromatogram - new All Signals Result table to display results for all signals in chromatogram together in one table.
*New: Chromatogram - New function Show Slope/Level that displays the values of the signal and signal derivation (slope) for purposes of filling the Fraction Table.
*New: Control: Agilent ICF libraries updated to version A.02.03 DU1.
*New: Control: All modules in development state or not validated as a part of official Clarity installation will be indicated by a yellow warning stripe in the Instrument window.
*New: Control: Alltesta Analyzer is available in testing state.
*New: Control: Column Usage Monitor - an auxiliary utility by UNI-RUBY script is available in testing state.
*New: Control: Ecom drivers updated ECD2000 to version 2.0.0.0, ECO2000 to 2.0.0.0, ECP2000 to 2.1.0.0., Flash06DAD to 1.6.0.0, Flash12DAD to 2.0.0.0.
*New: Control: Fraction Collector - change of color indicating collected fraction (from red to orange).
*New: Control: Fraction Collector - improved behavior and fixed bugs. It may be necessary to make amendments in existing scripts and methods.
*New: Control: Fraction Collector - option to collect on combination of Level and Slope values.
*New: Control: GL Science LC800 system - 64bit drivers added.
*New: Control: Gow Mac Series 816 GC control module has been released.
*New: Control: HTA HT2x00H - added support for two injectors.
*New: Control: HTA HT4000L autosampler control module has been released.
*New: Control: Knauer - PLATINblue Autosampler AS-1, UV Detector MW-1, Azura PDA Detector DAD2.1L, Azura UV Detector MWD2.1L control modules have been released.
*New: Control: Knauer control modules updated to version 6.0.0.5015.
*New: Control: Kontron 525 pump control module is available in Testing state.
*New: Control: Recipe HT4000 thermostat control module has been released.
*New: Control: Schambeck S 6300 autosampler control module has been released.
*New: Control: SSI Next Generation Pumps control module by UNI-RUBY script is available in testing state.
*New: Control: Sykam S1130 pump driver updated to version 1.0.12.0.
*New: Localization of OQ_Validation and LaunchManager utilities.
*New: Method Setup - for new methods, the External Start option default setting was changed from "Start/Restart" to Start Only.
*New: On stations connected to internet available update is indicated in the Main window status bar.
#Fixed Bug: Calibration - cloning - wrong date/time value has been used for the %variables.
#Fixed Bug: Chromatogram - Event table the "Input Run <" Type erroneously activated at start of run when the signal was higher than the set value (introduced in 5.0.0).
#Fixed Bug: Chromatogram - Export. Possible crashes during export of selected signal results from multiple signal chromatogram to Excel.
#Fixed Bug: Chromatogram - Export/Export Chromatogram did not remove invalid characters from signal name when only one signal was exported.
#Fixed Bug: Chromatogram - Groups selected in Result table were not highlighted in Chromatogram window (introduced in 6.0.0.295).
#Fixed Bug: Chromatogram - Menu Bar changes by Customize were lost upon closing the Chromatogram window (introduced in 6.1.0.130), other toolbars were not affected.
#Fixed Bug: Chromatogram - Corrected behavior for Set Initial colors in Chromatogram window graph properties.
#Fixed Bug: Control: Sedere ELSD models LC and FP - fixed problem at wake up from standby mode.
#Fixed Bug: Control: Shimadzu GC201x - dual inlet support for setting flow/pressure in Dual mode.
#Fixed Bug: Control: Shimadzu GC201x - PTV inlet wrong maximum heat rate limit.
#Fixed Bug: DataAcquisition - the Time and Signal range changes in one instrument were affecting also the Data Acquisition window opened on other Instruments. Introduced in 6.0.0.295
#Fixed Bug: DataAcquisition - fixed behavior for Set Initial Colors option.
#Fixed Bug: Extension GPC: the type of Instrument was incorrectly based on GC type instead of LC type.
#Fixed Bug: Extension MS: Under some rare circumstances a peak found on the quantification signal may appear twice in the Result table, once with method identification and once without.
#Fixed Bug: Extension SST: the limits set for RT column were not stored. Introduced in version 6.0.0.295, already fixed in version 6.1.0.130.
#Fixed Bug: OQ Validation - The limit for Height and Area %RSD for Peak 2 was adjusted.
#Fixed Bug: Reports - For dual signal chromatogram, the calibration curve and correlation were not printed in some situations.
#Fixed Bug: Sequence - problems with creating project subfolders from sequence (when a 2 or more level structure was attempted).